FDA Adverse Event Injury Summary report: N

NSK

MDR report key: 3889067 · Received May 9, 2014

Report

Report Number
1422375-2014-00001
Event Type
Injury
Date Received
May 9, 2014
Report Date
April 21, 2014
Manufacturer
NAKANISHI INC.
Product Code
KMW
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
GA, US
Reporter Occupation
DENTAL ASSISTANT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2014, NSK AMERICAN CORP. RECEIVED A REPAIR REQUEST FROM THE DENTAL OFFICE WITH THE DESCRIPTION/NEED FOR REPAIR DESCRIBED AS "BURNED PT. LIP." NO ADDITIONAL INFORMATION WAS PROVIDED. NSK AMERICAN CORP. CONTACTED THE DENTAL OFFICE ON (B)(6) 2014 IN AN ATTEMPT TO GATHER ADDITIONAL INFORMATION, SPECIFICALLY THE CONDITION OF THE PATIENT AND THE ACTION THAT WAS TAKEN AS A RESULT OF THE ALLEGED INCIDENT. THE DENTAL OFFICE PROVIDED THE FOLLOWING INFORMATION: AGE AND SEX OF THE PATIENT ((B)(6) YEAR OLD FEMALE); PROCEDURE BEING PERFORMED AT THE TIME OF THE ALLEGED INCIDENT AS THE EXTRACTION OF 4 WISDOM TEETH UNDER GENERAL/LOCAL ANESTHESIA; PRIOR TO THE ALLEGED INCIDENT, THERE WERE NO INDICATIONS THAT THE DEVICE HAD MALFUNCTIONED OR WAS NOT PERFORMING AS INTENDED; FIRST INDICATION OF A POTENTIAL INJURY WAS AN OBSERVATION OF MISSING SKIN FROM THE PATIENT'S LIP. FURTHER OBSERVATION SUGGESTED THE INJURY WAS A 2ND OR 3RD DEGREE BURN. THE PATIENT WAS PRESCRIBED PAIN MEDICATION AND ORAL ANTIBIOTICS; THE PATIENT RETURNED TO THE DENTAL OFFICE THE FOLLOWING WEEK FOR A PRE-SCHEDULED FOLLOW-UP; NO ADDITIONAL TREATMENT WAS NEEDED AT THAT TIME. THE DEVICE WAS RETURNED TO (B)(4) (INITIAL IMPORTER) ON (B)(4) 2014 FOR "REPAIR" PER THE CUSTOMER'S REQUEST. INITIAL VISUAL INSPECTION OF THE DEVICE INDICATES THAT THE DEVICE WAS NOT PROPERLY MAINTAINED ACCORDING TO THE INSTRUCTIONS FOR USE. A PRELIMINARY REVIEW OF THE LABELING INDICATES ADDITIONAL CLARIFICATION, WARNINGS, AND CAUTIONS MAY BE NEEDED. THE REPORT HAS BEEN SUBMITTED TO THE ORIGINAL MANUFACTURER. CLARIFICATION OF DATE OF THE EVENT, FOLLOW UP APPOINTMENT AND ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE OFFICE. WE ATTEMPTED CONTACT BY PHONE TWICE, FOLLOWED BY TWO LETTERS SENT ON 04/24/2014 AND 04/29/2014. NO RESPONSE HAS BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281085 NSK SGS-E2S KMW NAKANISHI INC. SGS-E2S

Patients

Seq Age Sex Outcome Treatment
1 18 YR Required Intervention