FDA Adverse Event
Malfunction
Summary report: N
POSEY HOOK & LOOP ALARM BELT LONG
MDR report key: 3888383
·
Received April 16, 2014
Report
- Report Number
- 2020362-2014-00146
- Event Type
- Malfunction
- Date Received
- April 16, 2014
- Report Date
- January 31, 2014
- Manufacturer
- J. T. POSEY COMPANY
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
RESULTS: EVALUATION OF THE RETURNED PRODUCTS REVEALED THAT WHEN THE ALARM WAS TESTED WITH A KNOWN WORKING SENSOR BELT, THE ALARM PASSED FUNCTIONAL TESTS PERFORMED. HOWEVER, WHEN THE ALARM WAS TESTED WITH THE RETURNED SENSOR BELT, THE ALARM TURNED ON/OFF WHEN THE BELT WAS MOVED. NO PHYSICAL DAMAGE WAS OBSERVED. THE RETURNED SENSOR BELT WAS RETESTED WITH A KNOWN WORKING ALARM, BUT THE RESULTS REMAIN THE SAME. THE TESTS REVEAL THAT THE SENSOR WIRES ARE BROKEN. (B)(4).
Description of Event or Problem · 1
CUSTOMER UNABLE TO PROVIDE ANY OTHER INFO THAN THE ALARM IS BROKEN. THE DATE THE ISSUE WAS DISCOVERED IS UNKNOWN. NO PATIENT INCIDENT OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 233650 | POSEY HOOK & LOOP ALARM BELT LONG | KMI | J. T. POSEY COMPANY | 8372L | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | ALARM, MODEL # 8350, SN (B)(4) |