DUAL SPEED CEMENT INJECTION GUN
Report
- Report Number
- 0001811755-2014-02218
- Event Type
- Malfunction
- Date Received
- June 20, 2014
- Date of Event
- May 26, 2014
- Report Date
- May 26, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- JDZ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE RETURNED DEVICE WAS CONFIRMED TO BE PRODUCING METAL SHAVINGS. BASED ON THE MANUFACTURING ENGINEER¿S EVALUATION OF THE DEVICE, THE SOURCE OF THE METAL SHAVINGS WAS DETERMINED TO BE THE RAM BUSHINGS. THE DEVICE WAS REPAIRED AND RETURNED TO THE USER FACILITY.
A FOLLOW UP REPORT WILL BE FILED AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED.
IT WAS REPORTED THAT DURING INSPECTION AT THE USER FACILITY, THE DUAL SPEED CEMENT INJECTION GUN WAS PRODUCING METAL SHAVINGS. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
IT WAS REPORTED THAT DURING INSPECTION AT THE USER FACILITY, THE DUAL SPEED CEMENT INJECTION GUN WAS PRODUCING METAL SHAVINGS. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 362648 | DUAL SPEED CEMENT INJECTION GUN | MIXER, CEMENT, FOR CLINICAL USE | JDZ | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |