FDA Adverse Event Malfunction Summary report: N

DUAL SPEED CEMENT INJECTION GUN

MDR report key: 3887226 · Received June 20, 2014

Report

Report Number
0001811755-2014-02218
Event Type
Malfunction
Date Received
June 20, 2014
Date of Event
May 26, 2014
Report Date
May 26, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
JDZ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED DEVICE WAS CONFIRMED TO BE PRODUCING METAL SHAVINGS. BASED ON THE MANUFACTURING ENGINEER¿S EVALUATION OF THE DEVICE, THE SOURCE OF THE METAL SHAVINGS WAS DETERMINED TO BE THE RAM BUSHINGS. THE DEVICE WAS REPAIRED AND RETURNED TO THE USER FACILITY.

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INSPECTION AT THE USER FACILITY, THE DUAL SPEED CEMENT INJECTION GUN WAS PRODUCING METAL SHAVINGS. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INSPECTION AT THE USER FACILITY, THE DUAL SPEED CEMENT INJECTION GUN WAS PRODUCING METAL SHAVINGS. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.