FDA Adverse Event Malfunction Summary report: N

SITTER ELITE

MDR report key: 3886459 · Received April 10, 2014

Report

Report Number
2020362-2014-00133
Event Type
Malfunction
Date Received
April 10, 2014
Report Date
January 31, 2014
Manufacturer
J. T. POSEY COMPANY
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

RESULTS: EVALUATION OF THE RETURNED UNIT CONFIRMED THAT THE NURSE CALL RECEPTACLE IS BROKEN AND HAS A LOOSE CONNECTION. THE NURSE CALL LED FUNCTIONS WHEN TESTED, BUT IF THE NURSE CALL CABLE IS MOVED, THE NURSE CALL WILL TURN ON AND OFF.

Description of Event or Problem · 1

CUSTOMER REPORTED THE NURSE CALL OUTLET TO THE ALARM IS BROKEN. CUSTOMER REPORTED ISSUE WAS DISCOVERED DURING SET-UP BUT A DATE WAS NOT PROVIDED. NO PATIENT INCIDENT OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
216641 SITTER ELITE KMI J. T. POSEY COMPANY 8345 NA

Patients

Seq Age Sex Outcome Treatment
1 NA CHAIR SENSOR: PART # 8309, LOT # UNK