FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 SINGLE C

MDR report key: 3886275 · Received April 9, 2014

Report

Report Number
9615050-2014-02601
Event Type
Malfunction
Date Received
April 9, 2014
Date of Event
March 11, 2014
Report Date
March 11, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE ALARMED WITH A S321 MALFUNCTION ALARM DURING THE DELIVERY ACCURACY TEST. A REVIEW OF THE DEVICE HISTORY INDICATED S321 MALFUNCTION ALARM WITH A MALFUNCTION CODE: QEP COUNT HIGH. THE PROBABLE CAUSE OF THE S321 MALFUNCTION ALARM CODE WAS A BROKEN MOTOR ENCODER. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH A S321 (MOTOR ERROR-PMC, LEFT) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A REPORT OF A S321 MALFUNCTION ALARM CODE. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
215994 SYMBIQ 3.13 SINGLE C 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK