T.W. POWER SUPPLY
Report
- Report Number
- 2242352-2014-00423
- Event Type
- Malfunction
- Date Received
- April 8, 2014
- Date of Event
- March 12, 2014
- Report Date
- March 14, 2014
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- HQO
- PMA / PMN Number
- K043155
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. A VISUAL DETERMINED THAT THE POWER SUPPLY CORD WAS NOT RETURNED. A PRE-CAUTERY TEST WAS PERFORMED WITH A REFERENCE EXTENSION CABLE AND HEMOPRO. THE REFERENCE DEVICE PASSED THE PRE-CAUTERY TEST; IT PRODUCED STEAM AND HEAT DURING SEVERAL ACTIVATIONS AND SHUT OFF WHEN THE TOGGLE WAS RELEASED. WE DID NOT IDENTIFY ANY NON-CONFORMITIES. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. SINCE THIS IS AN OEM SUPPLIED DEVICE, A LOT HISTORY REVIEW IS NOT APPLICABLE. (B)(4).
THE HOSP REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE T.W. POWER SUPPLY WAS TESTED AND THE DEVICE DID NOT WORK. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212868 | T.W. POWER SUPPLY | POWER SUPPLY | HQO | MAQUET CARDIOVASCULAR, LLC | VH-3010 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |