FDA Adverse Event Malfunction Summary report: N

T.W. POWER SUPPLY

MDR report key: 3886247 · Received April 8, 2014

Report

Report Number
2242352-2014-00423
Event Type
Malfunction
Date Received
April 8, 2014
Date of Event
March 12, 2014
Report Date
March 14, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
HQO
PMA / PMN Number
K043155
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. A VISUAL DETERMINED THAT THE POWER SUPPLY CORD WAS NOT RETURNED. A PRE-CAUTERY TEST WAS PERFORMED WITH A REFERENCE EXTENSION CABLE AND HEMOPRO. THE REFERENCE DEVICE PASSED THE PRE-CAUTERY TEST; IT PRODUCED STEAM AND HEAT DURING SEVERAL ACTIVATIONS AND SHUT OFF WHEN THE TOGGLE WAS RELEASED. WE DID NOT IDENTIFY ANY NON-CONFORMITIES. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. SINCE THIS IS AN OEM SUPPLIED DEVICE, A LOT HISTORY REVIEW IS NOT APPLICABLE. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE T.W. POWER SUPPLY WAS TESTED AND THE DEVICE DID NOT WORK. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212868 T.W. POWER SUPPLY POWER SUPPLY HQO MAQUET CARDIOVASCULAR, LLC VH-3010 NA

Patients

Seq Age Sex Outcome Treatment
1 NI