HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-19532
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- May 25, 2014
- Report Date
- May 25, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 ALARM (AIR IN LINE) OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING DWELL 3 OF 5. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. THE HP STATED THAT THE SUPPLY BAGS WERE EMPTY AND THERE WAS SOLUTION IN THE HEATER BAG ONLY. THE HP INFORMED THE TECHNICAL SERVICE REPRESENTATIVE (TSR) THAT THERE WERE NO BAGS WHICH CAME UNDONE. THERE WAS NOTHING UNUSUAL FOUND DURING THE TROUBLESHOOTING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ALARM. THE TSR EXPLAINED THE ALARM AND INSTRUCTED THE HP TO CYCLE THE POWER ON THE HC TO CLEAR IT. THE HP WOULD COMPLETE THE THERAPY WITH A MANUAL BAG. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 361520 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | HOMECHOICE |