FDA Adverse Event
Malfunction
Summary report: N
530G INSULIN PUMP
MDR report key: 3885193
·
Received June 19, 2014
Report
- Report Number
- 3004209178-2014-85978
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- May 27, 2014
- Report Date
- June 3, 2014
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
UNIT PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT TEST, REWIND, BASIC, OCCLUSION, OCCLUSION, PRIME/A33, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT FUNCTIONED PROPERLY. HOWEVER, LOUD MOTOR NOTED DURING TESTING DUE TO FAULTY MOTOR. MOTOR PASSED MOTOR TEST. UNIT RECEIVED WITH CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
CUSTOMER STATES THAT DUE TO FAULTY MOTOR THE INSULIN PUMP IS MAKING LOUD NOISES WHILE REWINDING. CUSTOMER STATES THAT SHE HAS NOT EXPERIENCED ANY ALARMS. THE CUSTOMER WAS ADVISED TO PERFORM A DISPLACEMENT TEST WITHOUT RESERVOIR, AND IT FAILED. THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. THE LAST BLOOD GLUCOSE READING WAS 102 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360476 | 530G INSULIN PUMP | OZO | OZO | MMT-551NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |