FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 3885193 · Received June 19, 2014

Report

Report Number
3004209178-2014-85978
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
May 27, 2014
Report Date
June 3, 2014
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

UNIT PASSED FUNCTIONAL TEST INCLUDING DISPLACEMENT TEST, REWIND, BASIC, OCCLUSION, OCCLUSION, PRIME/A33, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT FUNCTIONED PROPERLY. HOWEVER, LOUD MOTOR NOTED DURING TESTING DUE TO FAULTY MOTOR. MOTOR PASSED MOTOR TEST. UNIT RECEIVED WITH CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER STATES THAT DUE TO FAULTY MOTOR THE INSULIN PUMP IS MAKING LOUD NOISES WHILE REWINDING. CUSTOMER STATES THAT SHE HAS NOT EXPERIENCED ANY ALARMS. THE CUSTOMER WAS ADVISED TO PERFORM A DISPLACEMENT TEST WITHOUT RESERVOIR, AND IT FAILED. THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. THE LAST BLOOD GLUCOSE READING WAS 102 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360476 530G INSULIN PUMP OZO OZO MMT-551NAL

Patients

Seq Age Sex Outcome Treatment
1 66 YR