HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-19510
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- April 28, 2014
- Report Date
- May 27, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. AN INTERNAL AND EXTERNAL INSPECTION WAS PERFORMED. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED AS A HIGH DRAIN ERROR 105 (NIGHT DRAIN CYCLE FIVE) ALARM. THIS INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE. THE ALARM OCCURRED ON (B)(6) 2014 AT 14:12:20..NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360060 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |