FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 3884823 · Received June 19, 2014

Report

Report Number
2938836-2014-12226
Event Type
Injury
Date Received
June 19, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FOLLOW-UP, HIGH PACING THRESHOLD AND LOSS OF SENSING WERE OBSERVED. THE LEAD WAS REPOSITIONED AND THE PATIENT WAS DOING WELL AFTER EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360168 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7120Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention