DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Report
- Report Number
- 2938836-2014-12193
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- May 7, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED: COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A PARTIAL LEAD MEASURING 42.5CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 25.4-25.7CM FROM THE CUT END OF THE RV SHOCK COIL, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE SILICONE INSULATION WAS INTACT AT THIS LOCATION. (B)(6).
NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A PARTIAL LEAD MEASURING 42.5CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 6.5-7.1CM FROM THE PROXIMAL END OF THE SVC SHOCK COIL, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE SILICONE INSULATION WAS INTACT AT THIS LOCATION.
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360155 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7120Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |