FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 3884788 · Received June 19, 2014

Report

Report Number
2938836-2014-12191
Event Type
Injury
Date Received
June 19, 2014
Date of Event
April 25, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED MULTIPLE INAPPROPRIATE SHOCKS. UPON INTERROGATION HIGH, OUT OF RANGE, PACING LEAD IMPEDANCE AND LEAD NOISE EPISODES WERE NOTED. THE LEAD WAS EXPLANTED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359824 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention