RIATA ST ACTIVE FIXATION
Report
- Report Number
- 2938836-2014-12216
- Event Type
- Injury
- Date Received
- June 19, 2014
- Date of Event
- May 12, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0458
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
A PARTIAL LEAD WAS RETURNED IN TWO SEGMENTS FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 25.6-26.1CM, 26.6-26.9CM, 26.3-26.9CM, AND 27.1-27.9CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 11.5-13.6CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. EXTERNAL INSULATION ABRASION WAS NOTED AT 35.7-36.1CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS ABRADED AT THIS LOCATION.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
NEW INFORMATION RECEIVED NOTES THAT HIGH PACING THRESHOLD WAS ALSO OBSERVED. THE LEAD WAS EXPLANTED AND REPLACED.
IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE CLINIC FOR ROUTINE FOLLOW UP. DURING INTERROGATION, ATRIAL AND RV OVERSENSING WERE OBSERVED. IT WAS STATED THAT THE NOISE APPEARS TO BE CAUSED BY THE LEADS RUBBING TOGETHER. LEAD REVISION WAS RECOMMENDED. THE LEAD REMAINS IMPLANTED. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360687 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7000/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Required Intervention | (B)(4) |