FDA Adverse Event Malfunction Summary report: N

ANGLED ACET INSERTR

MDR report key: 3884595 · Received June 19, 2014

Report

Report Number
1818910-2014-21546
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

PREVIOUSLY A CORRECTIVE ACTION WAS IDENTIFIED FOR THIS FAILURE MODE WHEREBY A DESIGN CHANGE WAS IMPLEMENTED AND ROUTED ON ECO-193427 (DWG-106984) IN 2007 IN ORDER TO STRENGTHEN THE LOCKING CATCH. THE DESIGN VERIFICATION WAS COMPLETED AND ROUTED ON DVA-101743-FDE AND CONCLUDED THAT THERE ARE NO OUTSTANDING ACTIONS IDENTIFIED. THIS PRODUCT WAS MANUFACTURED TO THE REVISED DESIGN HOWEVER NO TREND IS IDENTIFIED AT THE CURRENT TIME. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PLASTIC THUMB LOCK BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360277 ANGLED ACET INSERTR HIP INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. 1818910 PC30404001

Patients

Seq Age Sex Outcome Treatment
1 62 YR