ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-04154
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- June 3, 2014
- Report Date
- June 11, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE TISSUE PAD DAMAGED, MELTED, AND A SMALL PORTION WAS NOT RETURNED. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND IT ACTIVATED DURING FUNCTIONAL TESTING. PROBABLE CAUSES OF TISSUE PAD DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. AVOID ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO PREVENT DAMAGE TO THE TISSUE PAD.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE WHITE TISSUE PAD WAS PIECE MISSING. THE MISSING PIECE WAS RETRIEVED BY FORCEPS. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360982 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |