FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3884513 · Received June 19, 2014

Report

Report Number
2032227-2014-03120
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
May 27, 2014
Report Date
May 27, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

DOWN ARROW BUTTON DID NOT RESPOND DUE TO FLATTENED BUTTON DOME SWITCH. NO FROZEN SCREEN NOTED. NO MOISTURE DAMAGE ON ELECTRONICS NOTED. THE INSULIN PUMP RECEIVED WITH SCRATCHED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED BELT CLIP SLOT AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE PATIENT STATED THAT THE PUMP HAD A FROZEN SCREEN AND UNRESPONSIVE BUTTONS. THE PATIENT STATED THAT IT WAS RAINING OUTSIDE, AND HER PUMP WAS UNDERNEATH HER SHIRT. ADVISED THE PATIENT THAT THE PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361369 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAP

Patients

Seq Age Sex Outcome Treatment
1 28 YR