FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3884355 · Received June 19, 2014

Report

Report Number
2955842-2014-03773
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
March 11, 2014
Report Date
May 30, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THE INSTRUMENT'S PITCH CABLE TO BE FRAYED. THE CONDUCTOR WIRES WERE INTACT AND UNDAMAGED HOWEVER THE DRIVE CABLE WAS FRAYED. THE PITCH UP CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. FRAYED STRANDS WERE OBSERVED TO BE STICKING OUT AT THE WRIST OF THE INSTRUMENT. THE OTHER CABLES AT THE WRIST OF THE INSTRUMENT WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI RADICAL RETROPUBIC PROSTATECTOMY SURGICAL PROCEDURE, THE SURGEON FOUND THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT DID NOT WORK. THE NURSE SLIPPED OUT THE INSTRUMENT AND OBSERVED THAT THE WIRE WAS EXPOSED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361874 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 S10130109 150

Patients

Seq Age Sex Outcome Treatment
1