FDA Adverse Event Injury Summary report: N

6.0MM CARBIDE DRILL BIT STERILE

MDR report key: 3884335 · Received June 19, 2014

Report

Report Number
9612488-2014-10235
Event Type
Injury
Date Received
June 19, 2014
Date of Event
May 21, 2014
Report Date
May 26, 2014
Manufacturer
SYNTHES BETTLACH
Product Code
HTW
PMA / PMN Number
PEXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: A MANUFACTURING EVALUATION WAS COMPLETED: OUR VISUAL INSPECTION HAS SHOWN THAT THE TIP OF THE DRILL BIT IS BLUNT AND WORN OUT. THESE DRILL BITS ARE SINGLE USE DEVICES AND IT IS LIKELY THAT THE TIP GOT BLUNT DUE TO NORMAL WEAR AND TEAR. LOT 7728655 WAS MANUFACTURED IN JANUARY 2011 TO SPECIFICATION. NO PRODUCT FAULT COULD BE FOUND. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL COMMON DEVICE NAME; HWE. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. NO NONCONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED A SURGEON WAS PERFORMING A REVISION SURGERY OF A PLATE SYSTEM, FOR AN UNKNOWN REASON AND WAS UNABLE TO REMOVE THE SCREWS DUE TO DAMAGED SCREW RECESS. THE SURGEON FIRST ATTEMPTED TO USE EXTRACTION SCREWS, BUT WAS NOT SUCCESSFUL. IT WAS REPORTED AN EXTRACTION SCREW BROKE DUE TO STRONG RESISTANCE, REPORTED TO BE LIKE A COLD WELDING CONDITION. THEN A CARBIDE DRILL BIT WAS USED, BUT THE BROKEN EXTRACT SCREW WAS NOT ABLE TO BE REMOVED. THE SURGEON DECIDED TO LEAVE THE PLATE AND SCREW IN THE PATIENT AND CLOSED THE OPERATION. IT WAS REPORTED THIS RESULTED IN A SURGICAL DELAY OF 120 MINUTES. THIS IS REPORT 1 OF 6 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359920 6.0MM CARBIDE DRILL BIT STERILE BIT,DRILL HTW SYNTHES BETTLACH 7728655

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention