FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3884267 · Received May 20, 2014

Report

Report Number
1720753-2014-04326
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 29, 2014
Report Date
May 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE AND THE IMAGE INTENSIFIER POWER SUPPLY WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO DISPLAY FLUOROSCOPIC EXPOSURES AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298968 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1