HEARTMATE II SYSTEM CONTROLLER
Report
- Report Number
- 2916596-2014-00814
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- April 18, 2014
- Report Date
- April 21, 2014
- Manufacturer
- THORATEC CORP.
- Product Code
- DSQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE SYSTEM CONTROLLER WAS RETURNED AND RECEIVED BY THE MANUFACTURER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE DEVICE ANALYSIS IS COMPLETED. NO FURTHER INFO IS AVAILABLE AT THIS TIME.
THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). IT WAS REPORTED THAT THE PT CONTACTED THE VAD COORDINATOR STATING THAT SHE HAD BEEN ON BATTERIES FOR APPROXIMATELY 18 HRS AND THOUGHT IT WAS "WEIRD" SHE HAD NOT RECEIVED A LOW BATTERY ALARM (SHE NORMALLY GETS 15-16 HRS ON BATTERY PER HER REPORT). WHEN THE PT LOOKED AT HER SYSTEM CONTROLLER IT WAS BLANK. NOTHING, INCLUDING THE GREEN ARROWS WERE ILLUMINATED AND THE PT RECEIVED NO AUDITORY ALARMS. THE PT SWITCHED HER SYSTEM CONTROLLER AT THIS POINT AND SHE REPORTED FEELING WHAT SHE THOUGHT WAS A "HEART BURN" BUT LOOKING BACK STATED IT MAY HAVE BEEN CHEST PAIN. THE VAD COORD. ON CALL MET THE PT AND GAVE HER A NEW BACKUP SYSTEM CONTROLLER. IT WAS UNDETERMINED HOW LONG THE PUMP WAS OFF. A REVIEW OF THE SUBMITTED LOG FILE DATA SHOWED A LOW ALARM (WHILE CONNECTED TO BATTERIES, A POWER CABLE DISCONNECT, AND NO EXTERNAL POWER ALARM WHERE THE EMERGENCY BACKUP BATTERY WAS IN USE, AND THE PUMP DROPPED TO 8000 RPMS, THE AUTO LOW SETTING, DUE TO POWER SAVER MODE BEING ACTIVATED. THEN IT APPEARS THE INTERNAL CLOCK/DATE WAS RESET TO 01/01/2001 AS THE BACK BATTERY BEING UNINSTALLED, THE PUMP STOPPED FOR 6 SECONDS, THEN THE WHITE CABLE FROM THE POWER MODULE WAS CONNECTED. THERE WAS THEN A YELLOW WRENCH ADVISORY FOLLOWED BY A DRIVELINE DISCONNECTED ALARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 298627 | HEARTMATE II SYSTEM CONTROLLER | DSQ: LVAD SYSTEM CONTROLLER | DSQ | THORATEC CORP. | 106762 | 130918 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR | Other |