FDA Adverse Event Malfunction Summary report: N

MPS DISPOSABLE

MDR report key: 3884188 · Received May 20, 2014

Report

Report Number
1649914-2014-00025
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
May 1, 2014
Report Date
May 1, 2014
Manufacturer
QUEST MEDICAL, INC.
Product Code
DTR
PMA / PMN Number
K953838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). QUEST MEDICAL, INC. HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. QUEST MEDICAL, INC. DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE FOREIGN DISTRIBUTOR ((B)(6)) REPORTED A LEAKAGE ISSUE WITH THE MPS DISPOSABLE DELIVERY SET FROM ONE OF THEIR CUSTOMERS. THE HOSPITAL REPORTED A LEAKAGE ORIGINATING AT THE CONNECTION BETWEEN THE "MEDICAL SOLUTION BAG" (EITHER ARREST AGENT CASSETTE OR ADDITIVE CASSETTE) AND THE HEAT EXCHANGER. THE CUSTOMER REPORTED TIGHTENING THE CONNECTOR AGAIN BUT THE ALLEGED LEAK PERSISTED. IT WAS REPORTED THAT THE DISPOSABLE SET WAS USED FOR THE ENTIRE PROCEDURE WITH NO IMPACT TO THE PATIENT. THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE DEVICE LOT NUMBER WAS NOT RECORDED BY THE CUSTOMER AND IS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298697 MPS DISPOSABLE CPBP HEAT EXCHANGER DTR QUEST MEDICAL, INC. 5001102J

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention