FDA Adverse Event
Malfunction
Summary report: N
CARE ASSIST BED
MDR report key: 3883956
·
Received February 14, 2014
Report
- Report Number
- 9615739-2014-00042
- Event Type
- Malfunction
- Date Received
- February 14, 2014
- Date of Event
- January 16, 2014
- Report Date
- January 16, 2014
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE BRAKE CASTER AND BRAKE STEER CASTERS ARE WORN. THE TECHNICIAN REPLACED THE BRAKE CASTER AND BRAKE STEER CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
THE ACCOUNT REPORTED THE BRAKES ARE NOT HOLDING. THE BED IS LOCATED ON THE SECOND FLOOR AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 98524 | CARE ASSIST BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |