FDA Adverse Event Malfunction Summary report: N

AED PLUS

MDR report key: 3883884 · Received February 13, 2014

Report

Report Number
1220908-2014-00328
Event Type
Malfunction
Date Received
February 13, 2014
Report Date
January 29, 2014
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K011541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY RESISTOR ON THE MAIN BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95676 AED PLUS DEFIBRILLATOR MKJ ZOLL MEDICAL CORPORATION AED PLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA