FDA Adverse Event
Malfunction
Summary report: N
REMAN TC BARIATRIC BED
MDR report key: 3883851
·
Received February 13, 2014
Report
- Report Number
- 1824206-2014-00415
- Event Type
- Malfunction
- Date Received
- February 13, 2014
- Date of Event
- January 16, 2014
- Report Date
- January 16, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- PMA / PMN Number
- K122473
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND ONE OF THE REAR THE CASTERS NEEDING TO BE REPLACED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. CASTER WAS REPLACED AND THE ISSUE WAS RESOLVED. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM SERVICE TECHNICIAN STATING, "BRAKES SETTING, BUT WILL NOT HOLD." THERE WAS NO PATIENT/USER INJURY REPORTED. T(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96797 | REMAN TC BARIATRIC BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | PR1840A000001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |