FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC HEAT PACK

MDR report key: 3883847 · Received June 13, 2014

Report

Report Number
1811605-2014-00133
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 15, 2014
Report Date
June 13, 2014
Manufacturer
BATTLE CREEK EQUIPMENT
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

OUR EXAMINATION REVEALED THAT ONE OF THE TERMINALS HAD BEEN BROKEN NEAR A THERMOSTAT, WHICH HAD COMPROMISED OUR DOUBLE-INSULATED DESIGN. WE ALSO FOUND SEVERAL OTHER INTERNAL COMPONENTS THAT WERE DAMAGED, INDICATING THAT THE CUSTOMER MISUSED THE PAD BY APPLYING AN EXCESSIVE, LOCALIZED FORCE ON OR NEAR THE THERMOSTAT TERMINAL. WE CONCLUDED THAT THIS REPORT WAS ATTRIBUTED TO MISUSE BY THE USER, AND FAILURE TO FOLLOW INSTRUCTIONS AS TO THE CARE AND HANDLING OF THE UNIT. ALTHOUGH WARNING LABELS AND INSTRUCTION BOOKLETS WARN AGAINST SITTING OR LAYING ON THE PRODUCT, WE HAVE A CORRECTIVE ACTION PROJECT (B)(4) UNDERWAY TO MAKE THE DESIGN MORE ROBUST.

Description of Event or Problem · 1

WHILE SPEAKING WITH OUR C.S., THE CUSTOMER PLUGGED THE PAD IN AND IT CAUGHT FIRE.