FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC HEAT PACK
MDR report key: 3883847
·
Received June 13, 2014
Report
- Report Number
- 1811605-2014-00133
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- May 15, 2014
- Report Date
- June 13, 2014
- Manufacturer
- BATTLE CREEK EQUIPMENT
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
OUR EXAMINATION REVEALED THAT ONE OF THE TERMINALS HAD BEEN BROKEN NEAR A THERMOSTAT, WHICH HAD COMPROMISED OUR DOUBLE-INSULATED DESIGN. WE ALSO FOUND SEVERAL OTHER INTERNAL COMPONENTS THAT WERE DAMAGED, INDICATING THAT THE CUSTOMER MISUSED THE PAD BY APPLYING AN EXCESSIVE, LOCALIZED FORCE ON OR NEAR THE THERMOSTAT TERMINAL. WE CONCLUDED THAT THIS REPORT WAS ATTRIBUTED TO MISUSE BY THE USER, AND FAILURE TO FOLLOW INSTRUCTIONS AS TO THE CARE AND HANDLING OF THE UNIT. ALTHOUGH WARNING LABELS AND INSTRUCTION BOOKLETS WARN AGAINST SITTING OR LAYING ON THE PRODUCT, WE HAVE A CORRECTIVE ACTION PROJECT (B)(4) UNDERWAY TO MAKE THE DESIGN MORE ROBUST.
Description of Event or Problem · 1
WHILE SPEAKING WITH OUR C.S., THE CUSTOMER PLUGGED THE PAD IN AND IT CAUGHT FIRE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 351377 | THERMOPHORE AUTOMATIC HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT | 155 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |