FDA Adverse Event Malfunction Summary report: N

TOTAL CARE

MDR report key: 3883840 · Received February 13, 2014

Report

Report Number
1824206-2014-00421
Event Type
Malfunction
Date Received
February 13, 2014
Date of Event
January 16, 2014
Report Date
January 16, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM SERVICE TECHNICIAN INSPECTED THE BED AND FOUND THAT THE STEER/BRAKE CASTER WAS RATCHETING AND WOULD NOT HOLD. HE REPLACED THE STEER/BRAKE CASTER AND THE ISSUE WAS RESOLVED. THE MOST LIKELY CAUSE FOR THIS FAILURE IS WEAR. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN ONCE IN 2012 AND ONCE IN 2013.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM HILL-ROM SERVICE TECHNICIAN STATING, "THE STEER CASTER WAS RATCHETING AND WOULD NOT HOLD." THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96734 TOTAL CARE A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1900D003346