FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 3883831 · Received February 13, 2014

Report

Report Number
1824206-2014-00417
Event Type
Malfunction
Date Received
February 13, 2014
Date of Event
January 16, 2014
Report Date
January 16, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE SIDECOM COMMUNICATION CABLE WAS NOT CONNECTED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BEDS. THE TECHNICIAN RECONNECTED THE SIDECOM COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE NURSE CALL WOULD NOT WORK. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96733 VERSACARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1