FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 3883727
·
Received June 19, 2014
Report
- Report Number
- 0001811755-2014-02205
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- May 26, 2014
- Report Date
- May 26, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN WAS UNABLE TO DUPLICATE THE REPORTED EVENT OF HEATING UP BUT NOTED THAT THE DRIVESHAFT BEARING WAS DAMAGED.
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACK-UP EQUIPMENT WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACK-UP EQUIPMENT WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360100 | CORE IMPACTION DRILL | DRIVER, WIRE, AND BONE DRILL, MANUAL | DZJ | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |