HEARTWARE® VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2014-00637
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- March 13, 2014
- Report Date
- May 21, 2014
- Manufacturer
- HEARTWARE, INC
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1607-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THE BATTERY WAS RETURNED; VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICE IN RELATION TO THE REPORTED EVENT. THOROUGH EXTERNAL VISUAL INSPECTION OF THE BATTERY REVEALED NO SIGNS OF PHYSICAL DAMAGE OR CONTAMINATION. FUNCTIONAL TESTING REVEALED THE RETURNED BATTERY CONTAINED A FAULTY CELL PAIR, WHICH MOST LIKELY CONTRIBUTED TO THE REPORTED EVENT. AN INTERNAL INVESTIGATION (CAPA) HAS BEEN OPENED TO ADDRESS THE ISSUE.
THE DEVICE HAS BEEN RECEIVED AND EVALUATION STILL ONGOING. PRELIMINARY EVALUATION AND TESTING OF THE RETURNED BATTERY REVEALED AN INTERNAL FAULTY CELL. THIS FINDING WAS RE-ASSESSED PER OUR SOP REQUIREMENTS, AND DETERMINED TO BE REPORTABLE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF RECEIPT. EVALUATION IN PROGRESS.
APPROXIMATELY ONE YEAR AND SIX MONTHS POST HVAD IMPLANTATION, THE SITE REPORTED THAT THE PATIENT EXPERIENCED POWER SOURCE CHANGE IN THE CONTROLLER EARLIER THAN EXPECTED. THE BATTERY WAS REMOVED FROM THE PATIENT AND A NEW BATTERY WAS SUPPLIED. NO HARM OR INJURY TO THE PATIENT WAS REPORTED AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360069 | HEARTWARE® VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM, BATTERY | DSQ | HEARTWARE, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4)- BATTERY |