FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3883703 · Received June 19, 2014

Report

Report Number
2032227-2014-03082
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
May 26, 2014
Report Date
May 26, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP HAD MISSING SEGMENTS/PARTIAL DISPLAY DUE TO CRACKED ZEBRA CONNECTOR ON LCD. THE INSULIN PUMP HAD BROKEN BATTERY TUBE THREADS, BROKEN RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD HIGH BLOOD GLUCOSE OF 520 MG/DL, AND HER INSULIN PUMP WAS MALFUNCTIONING. CALLER STATED THAT THE CUSTOMER'S INSULIN PUMP WAS ALARMING, SQUIRTING OUT THE INSULIN DURING MANUAL PRIME, WAS UNABLE TO EXIT FROM PREPARING TO PRIME LOOP AND THE DRIVE SUPPORT CAP IS STICKING OUT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360035 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAS

Patients

Seq Age Sex Outcome Treatment
1 70 YR