CAPTIFLEX?
Report
- Report Number
- 3005099803-2014-02279
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Report Date
- May 28, 2014
- Manufacturer
- BOSTON SCIENTIFIC - COSTA RICA (COYOL)
- Product Code
- FDI
- PMA / PMN Number
- K941750
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
VISUAL EVALUATION OF THE RETURNED DEVICE FOUND THE WIRE KINKED NEAR TO THE PROXIMAL SECTION AND THE HANDLE CANNULA DETACHED. THE HANDLE CANNULA HAD MARKS FROM THE HANDLE ASSEMBLY PROCESS PERFORMED DURING THE MANUFACTURING. ALSO, THE INTERNAL SURFACE OF THE FINGER RING WAS CHECKED AND ONE EDGE OF THE INTERNAL GROOVE IS DAMAGED. FUNCTIONAL EVALUATION COULD NOT BE PERFORMED DUE TO THE CONDITION OF THE RETURNED DEVICE. THE COMPLAINT WAS CONFIRMED; THE HANDLE CANNULA WAS DETACHED. HOWEVER, REVIEW AND ANALYSIS OF ALL AVAILABLE INFORMATION FAILED TO INDICATE A ROOT CAUSE OR PROBABLE ROOT CAUSE FOR THIS EVENT. THERE IS AN INVESTIGATION IN PLACE TO ADDRESS THIS ISSUE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS AT THE TIME OF RELEASE TO DISTRIBUTION.
REPORTED EVENT OF HANDLE CANNULA DETACHED. THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS; HOWEVER, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CAPTIFLEX SMALL OVAL SNARE WAS USED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE HANDLE CANNULA DETACHED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CAPTIFLEX SMALL OVAL SNARE WAS USED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE HANDLE CANNULA DETACHED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359781 | CAPTIFLEX? | SNARE, FLEXIBLE | FDI | BOSTON SCIENTIFIC - COSTA RICA (COYOL) | M00562422 | 0016744770 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |