OPTICROSS?
Report
- Report Number
- 2134265-2014-03415
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- March 5, 2014
- Report Date
- May 22, 2014
- Manufacturer
- BOSTON SCIENTIFIC - FREMONT (SUD)
- Product Code
- OBJ
- PMA / PMN Number
- K123621
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE RETURNED TO MFR: THE DEVICE WAS RETURNED FOR EVALUATION. EXAMINATION OF THE RETURNED DEVICE REVEALED KINKS WERE OBSERVED IN THE TELESCOPE ASSEMBLY AT 71.6CM AND 67.3CM FROM FEMORAL MARKER TO THE PROXIMAL END. THE TELESCOPE ASSEMBLY WAS NOT ABLE TO PROPERLY PULL BACK, ADVANCE, OR RETRACT. THE SHEATH ASSEMBLY WAS FOUND TO HAVE THE CLEAR TUBING DETACHED FROM THE BLUE TUBING AT THE LAP JOINT. THIS FULLY EXPOSES THE IMAGING CORE. SINCE THE CLEAR TUBING WAS DETACHED, THE DISTANCE FROM THE DISTAL END OF THE TRANSDUCER HOUSING TO THE TIP OF THE CATHETER WAS NOT MEASURED. DURING IMAGE CHARACTERIZATION TESTING, NO IMAGE APPEARED IN THE SYSTEM DUE TO ELECTRICAL OPEN AT PROXIMAL. IN ORDER TO INSPECT IMAGING CORE WINDUP AT THE PROXIMAL END OF THE CATHETER, THE RETAINER CLIP WAS REMOVED AND IMAGING CORE ASSEMBLY WAS PULLED OUT FROM HUB. IMAGING CORE WINDUP WAS FOUND IN THE TELESCOPE ASSEMBLY. NO OTHER ISSUES OR DEFECTS WERE OBSERVED DURING PRODUCT ANALYSIS OF THE RETURNED DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(4) 2014. IT WAS REPORTED THAT THE DRIVE SHAFT WAS STIFF AND DID NOT MOVE BACK TO THE ITS ORIGINAL POSITION. THE 90% STENOSED, MODERATELY CALCEFIED AND MODERATELY TORTUOUS LESION WAS LOCATED IN THE MIDDLE PORTION OF THE LEFT ANTERIOR DESCENDING ARTERY (LAD). AN OPTICROSS¿WAS USED TO VIEW THE LESION. THE DEVICE WAS USED WITHOUT ANY ISSUES DURING SET UP AND IMAGING BUT AFTER THEY ATTEMPTED AUTOMATIC PULLBACK IT WAS REPORTED THAT THE DRIVE SHAFT WAS STIFF AND IT WOULD NOT MOVE BACK TO ITS INITIAL POSITION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS IS GOOD. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE SHEATH ASSEMBLY DETACHED FROM THE BLUE TUBING AT THE LAP JOINT AREA OF THE CATHETER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 361132 | OPTICROSS? | CATHETER, ULTRASOUND, INTRAVASCULAR | OBJ | BOSTON SCIENTIFIC - FREMONT (SUD) | H749518080 | 16484831 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |