ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-17604
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Report Date
- June 14, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 07/17/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/08/2014 WITH THE FOLLOWING FINDINGS:A VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE KEYPAD COVER WAS INTACT. THE PUMP POWERED UP WITH THE APPROPRIATE AUDIBLE AND VIBRATORY FEATURES. ALL THE KEYPAD BUTTONS RESPONDED INTERMITTENTLY. REMOVAL OF THE KEYPAD COVER FOUND CONTAMINATION UNDER ALL OF THE BUTTON CONTACTS. UNRELATED TO THE BUTTON ISSUE, IT WAS OBSERVED THAT THE VERIFY SCREEN WAS DIM AND DISCOLORED AND THE BATTERY CAP WAS DAMAGED.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. REPORTEDLY, THE ¿UP¿, ¿DOWN¿ AND ¿OK¿ BUTTONS WERE UNDER RESPONSIVE. IT WAS REPORTED THAT THE KEYPAD COVER WAS UNDAMAGED AND THERE WAS NO EVIDENCE OF MOISTURE OR CORROSION IN THE PUMP. THE REPORTER DID NOT NOTE ANY DELIVERY ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE, THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 361523 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |