FDA Adverse Event Malfunction Summary report: N

ECHELON LINEAR CUTTER

MDR report key: 3883433 · Received June 19, 2014

Report

Report Number
3005075853-2014-04113
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
April 17, 2014
Report Date
April 29, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: NICKED KNIFE. THE ANALYSIS FOUND THAT THE LONG60A DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITHOUT A RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND IT FIRED, AND FORMED ALL THE STAPLES AS INTENDED. THE CUT WAS NOTED TO BE JAGGED DUE TO A DAMAGED KNIFE. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE TO THE KNIFE IS WHEN THE DEVICE IS FIRED OVER AN ALREADY EXISTING STAPLE LINE, HARD OBJECT OR THICKER TISSUE THAN INDICATED. REPEATEDLY FIRING ACROSS EXISTING STAPLE LINES CAN ALSO REDUCE THE ABILITY TO CUT CLEANLY. TO MITIGATE THE POTENTIAL FOR STAPLES GETTING INTO THE CARTRIDGE AND INTERFERING WITH THE KNIFE PATH DURING DEVICE FIRING, PRIOR TO RELOADING THE DEVICE, RINSE THE ANVIL AND CARTRIDGE JAW IN STERILE SOLUTION AND THEN WIPE THE ANVIL AND CARTRIDGE JAW TO CLEAN ANY FORMED BUT UNUSED STAPLES FROM THE DEVICE. ADDITIONALLY, PROPER CARE SHOULD BE TAKEN WHEN PLACING THE DEVICE ON THE TISSUE TO BE STAPLED, TO ENSURE THAT NO HARD OBSTRUCTION SUCH AS A CLIP IS INCLUDED WITH THE TISSUE INSIDE THE JAWS. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SLEEVE GASTRECTOMY, THERE WAS ¿STIFF FIRING¿ OF THE DEVICE. OPENED UP LIKE DEVICE AND FINISHED PROCEDURE SUCCESSFULLY. 30-SECOND DELAY IN PROCEDURE. NO ADVERSE EVENT TO PATIENT. PATIENT DETAILS NOT PROVIDED. USER IS TRAINED AND DEVICE WAS STORED TO LABELING INSTRUCTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361474 ECHELON LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA K4DG32

Patients

Seq Age Sex Outcome Treatment
1