FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3883395 · Received June 19, 2014

Report

Report Number
3004209178-2014-85681
Event Type
Injury
Date Received
June 19, 2014
Date of Event
June 13, 2014
Report Date
June 16, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, SELF TEST AND THE ERROR TEST. ALL OPERATING CURRENTS ARE WITHIN SPECIFICATION. OFF NO POWER ALARM FUNCTIONS PROPERLY. THE DEVICE WAS UNABLE TO PRIME DURING PRIME TEST AND ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO FAULTY FORCE SENSOR. UNABLE TO PERFORM THE OCCLUSION TEST AND EXCESSIVE NO DELIVERY TEST DUE TO PRIME/FILL ANOMALY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE UNIT HAD MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BELT CLIP SLOT, SCRATCHED RESERVOIR TUBE WINDOW AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER'S MOTHER STATED THAT THE INSULIN PUMP ALARM MOTOR ERROR THREE TIMES. CALLER STATED THAT CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE. CUSTOMER WAS VOMITING AND KETONES. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361002 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWH

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization