PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-85681
- Event Type
- Injury
- Date Received
- June 19, 2014
- Date of Event
- June 13, 2014
- Report Date
- June 16, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, SELF TEST AND THE ERROR TEST. ALL OPERATING CURRENTS ARE WITHIN SPECIFICATION. OFF NO POWER ALARM FUNCTIONS PROPERLY. THE DEVICE WAS UNABLE TO PRIME DURING PRIME TEST AND ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO FAULTY FORCE SENSOR. UNABLE TO PERFORM THE OCCLUSION TEST AND EXCESSIVE NO DELIVERY TEST DUE TO PRIME/FILL ANOMALY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE UNIT HAD MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BELT CLIP SLOT, SCRATCHED RESERVOIR TUBE WINDOW AND CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER'S MOTHER STATED THAT THE INSULIN PUMP ALARM MOTOR ERROR THREE TIMES. CALLER STATED THAT CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE. CUSTOMER WAS VOMITING AND KETONES. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 361002 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722WWH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |