FDA Adverse Event Malfunction Summary report: N

HIGH DEFINITION LCD MONITOR

MDR report key: 3883305 · Received February 20, 2014

Report

Report Number
8010047-2014-00058
Event Type
Malfunction
Date Received
February 20, 2014
Date of Event
January 23, 2014
Report Date
January 24, 2014
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
NWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AFTER THE PROCEDURE, THE SUBJECT DEVICE WAS INSPECTED BY THE MEDICAL ENGINEER OF THE FACILITY. THE IMAGE OF THE SUBJECT DEVICE APPEARED AT THAT TIME, BUT IT WAS PLANNED TO RETURN TO OLYMPUS FOR INSPECTION AND REPAIR. THERE WAS NO IRREGULARITY OF THE SUBJECT DEVICE WITH THE DEVICE HISTORY RECORD. BECAUSE OF THE HOLE IN THE SALINE PACK WHICH WAS MADE BY THE IMPROPER HANDLING OF THE FACILITY, THE ELECTRICAL CIRCUIT INSIDE THE SUBJECT DEVICE BECAME WET WITH THE SALINE, BECAME SHORT STATE, AND EVENTUALLY THE SUBJECT DEVICE BECAME NOT FUNCTIONING PROPERLY. SEE SCANNED PAGE. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP (OMSC) WAS INFORMED THAT DURING AN ESOPHAGEAL TUMOR RESECTION WITH THE SUBJECT DEVICE (OEV261H) AND OTHER DEVICES SET ON A WORKING STATION, WHEN A SALINE PACK WAS PRESSURIZED AND HANG ON THE WORKING STATION'S HANGER FOR CABLES, A HOLE WAS MADE ON THE PACK AND SALINE WAS POURED ON THE SUBJECT DEVICE AND OTHER DEVICES. IT CAUSED THE DISAPPEARANCE OF IMAGE ON THE MONITOR OF THE SUBJECT DEVICE. THE SET OF THE WORKING STATION INCLUDING THE SUBJECT DEVICE WAS REPLACED TO ANOTHER SET OF WORKING STATION, AND THE PROCEDURE WAS CONTINUED AND COMPLETED. THERE WAS NO PT OR MEDICAL STAFF INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106809 HIGH DEFINITION LCD MONITOR LCD MONITOR NWB OLYMPUS MEDICAL SYSTEMS CORPORATION OEV261H NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Other