FDA Adverse Event Malfunction Summary report: N

AFFINITY THREE BIRTHING BED

MDR report key: 3883281 · Received February 20, 2014

Report

Report Number
1824206-2014-00532
Event Type
Malfunction
Date Received
February 20, 2014
Date of Event
January 23, 2014
Report Date
January 23, 2014
Manufacturer
HILL-ROM INC.
Product Code
KNC
PMA / PMN Number
K915779
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT FOUND THE BRAKE CASTER AND BRAKE STEER CASTER ROTATE TO THE SIDE. PER THE HILL-ROM SERVICE MANUAL, THE AFFINITY THREE BIRTHING BED AND AFFINITY FOUR BIRTHING BEDS REQUIRE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM SEMI-ANNUAL PREVENTATIVE MAINTENANCE. CHECK THE TIRES FOR CUTS, WEAR, TREAD LIFE, ETC. APPLY THE BRAKE, AND CHECK TO ENSURE THAT THE BED WILL NOT MOVE. IF THE BED MOVES, INSPECT IT FOR WEAR, AND ADJUST IF REQUIRED. APPLY THE STEERING PEDAL AND CHECK THE STEERING TO ENSURE PROPER LOCKING ACTION WHEN ACTIVATED. THE ACCOUNT REPLACED THE BRAKE CASTER AND BRAKE STEER CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BRAKE CASTER AND BRAKE STEER CASTERS DO NOT HOLD WHILE IN BRAKE MODE. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107192 AFFINITY THREE BIRTHING BED BIRTHING BED KNC HILL-ROM INC. 3700

Patients

Seq Age Sex Outcome Treatment
1