SYMBIQ SINGLE CHANNEL
Report
- Report Number
- 9615050-2014-03584
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- January 1, 2014
- Report Date
- April 23, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) ALARM CODE. THE PROBABLE CAUSE OF THE S233 MALFUNCTION ALARM CODE WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO INFO WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, A S233 MALFUNCTION ALARM CODE WAS NOTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309051 | SYMBIQ SINGLE CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |