FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ DUAL CHANNEL
MDR report key: 3883064
·
Received May 23, 2014
Report
- Report Number
- 9615050-2014-03619
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- March 26, 2014
- Report Date
- April 25, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0070-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE DEVICE TOUCHSCREEN WAS OUT OF ALIGNMENT. DURING REVIEW OF THE DEVICE HISTORY MULTIPLE BUTTON ID: INVALID ERROR CODES WERE NOTED. FURTHER TESTING FOUND CONTAMINATION ON THE BOTTOM TOUCHSCREEN CONNECTIONS DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN WAS UNABLE TO CALIBRATE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307545 | SYMBIQ DUAL CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |