FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3882796
·
Received April 10, 2014
Report
- Report Number
- 1720753-2014-03116
- Event Type
- Malfunction
- Date Received
- April 10, 2014
- Date of Event
- March 24, 2014
- Report Date
- April 10, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE WAS EVALUATED AND REPLACED. THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM LOST PATIENT DATA AND IMAGES. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 216495 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |