FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 3882743
·
Received April 10, 2014
Report
- Report Number
- 1720753-2014-03143
- Event Type
- Malfunction
- Date Received
- April 10, 2014
- Date of Event
- March 24, 2014
- Report Date
- April 10, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE CPU COIN BATTERY WAS EVALUATED AND REPLACED. THE CMOMS SETTINGS WERE ALSO RELOADED AS PART OF THE SERVICE EVENT. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT TO A USABLE STATE AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 216485 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |