FDA Adverse Event Injury Summary report: N

M2A 38MM MOD HD+6MM NK NO SKRT

MDR report key: 3882707 · Received June 19, 2014

Report

Report Number
0001825034-2014-05605
Event Type
Injury
Date Received
June 19, 2014
Date of Event
May 23, 2014
Report Date
November 6, 2014
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK011110
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT IDENTIFICATION NECESSARY TO REVIEW MANUFACTURING HISTORY WAS NOT PROVIDED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. DEVICE AVAILABILITY - THE DEVICE IS REPORTED TO BE AVAILABLE FOR EVALUATION; HOWEVER, IT HAS NOT BEEN RECEIVED BY BIOMET ORTHOPEDICS TO DATE. IN THE EVENT THAT THE DEVICE IS RECEIVED AND EVALUATED, A FOLLOW UP REPORT WILL BE SENT TO THE FDA TO PROVIDE RESULTS.

Additional Manufacturer Narrative · 1

EXAMINATION OF RETURNED DEVICE FOUND NO EVIDENCE OF PRODUCT NON-CONFORMANCE. DURING THE MATERIAL ANALYSIS, IT WAS NOTED THE COMPONENT SHOWED EVIDENCE OF SUBLUXATION OF THE JOINT, SCRATCHING OF THE BEARING SURFACES AND MINOR TAPER FRETTING.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. EVALUATION IN PROCESS BUT NOT YET COMPLETE. UPON COMPLETION OF EVALUATION, A FOLLOW UP REPORT WILL BE SENT TO THE FDA. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ THIS REPORT IS NUMBER 1 OF 2 MDR'S FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-05605 AND 08359).

Description of Event or Problem · 1

IT WAS REPORTED PATIENT UNDERWENT A TOTAL HIP ARTHROPLASTY ON (B)(6) 2005. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2014 ALLEGEDLY DUE TO ADVERSE REACTION TO METAL DEBRIS AND PERSONAL INJURY. THE MODULAR HEAD AND ACETABULAR CUP WERE REMOVED AND REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED PATIENT UNDERWENT A TOTAL HIP ARTHROPLASTY ON (B)(6) 2005. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2014 ALLEGEDLY DUE TO ADVERSE REACTION TO METAL DEBRIS AND PERSONAL INJURY. THE MODULAR HEAD AND ACETABULAR CUP WERE REMOVED AND REPLACED. ADDITIONAL INFORMATION REVEALED PRODUCT IDENTIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360573 M2A 38MM MOD HD+6MM NK NO SKRT PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 908470

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R