FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3882694 · Received June 19, 2014

Report

Report Number
2939301-2014-14904
Event Type
Malfunction
Date Received
June 19, 2014
Report Date
June 11, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (11/04/2014). THE PATIENT¿S METER HAS BEEN RETURNED ON 9/24/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 10/22/2014 WITH THE FOLLOWING FINDINGS:THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE A DEFECTIVE CAPACITOR C85. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR 1. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361159 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3522706

Patients

Seq Age Sex Outcome Treatment
1 69 YR