FDA Adverse Event Malfunction Summary report: N

SYMBIQ SCHANNEL 3.01

MDR report key: 3882680 · Received April 10, 2014

Report

Report Number
9615050-2014-02584
Event Type
Malfunction
Date Received
April 10, 2014
Date of Event
March 11, 2014
Report Date
March 11, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
FA212-03
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE PASSED TESTING. ALTHOUGH THE DEVICE PASSED TESTING, A REVIEW OF THE HISTORY INDICATED A WHITESCREEN ERROR. THE WHITESCREEN ERROR OCCURRED WHEN THE USER TITRATED ONLY THE RATE WHEN THE DEVICE WAS DELIVERING IN THE KVO MODE. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.