COULTER LH 500 HEMATOLOGY ANALYZER

MDR report key: 3882421 · Received June 18, 2014

Report

Report Number
1061932-2014-01366
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
May 25, 2014
Report Date
May 25, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE OBSERVED A LEAK DUE TO PINHOLE IN THE TUBING GOING THROUGH PINCH VALVE (PV43). THE FSE REPLACED THE TUBING AND VERIFIED THE INSTRUMENT. THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY 2 ML OF FLUID LEAKED FROM THE COULTER LH 500 HEMATOLOGY ANALYZER. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THERE WAS NO REPORTED EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE CUSTOMER WAS WEARING GLOVES AND A LAB COAT AT THE TIME OF THE INCIDENT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION WITH THE REPORTED EVENT. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
357197 COULTER LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA