COULTER LH 780 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2014-01379
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 21, 2014
- Report Date
- May 21, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061616
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE OBSERVED A LEAK DUE TO THE QUICK DISCONNECTS ON THE TUBING THROUGH PINCH VALVES VL46B AND VL41 WERE LOOSE. THE FSE TIGHTENED THE QUICK DISCONNECTS AND THE INSTRUMENT RAN WITHOUT ANY LEAKS. THE FSE VERIFIED THE INSTRUMENT AND THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).
THE CUSTOMER REPORTED APPROXIMATELY HALF A CUP OF CLEAR FLUID LEAKED INSIDE THE COULTER LH 780 HEMATOLOGY ANALYZER. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AT THE TIME OF THE OCCURRENCE. THERE WAS NO EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. NO ERRONEOUS PATIENT RESULTS WERE GENERATED IN CONNECTION WITH THIS EVENT. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 356870 | COULTER LH 780 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |