FDA Adverse Event Malfunction Summary report: N

COULTER LH 500 HEMATOLOGY ANALYZER

MDR report key: 3882391 · Received June 18, 2014

Report

Report Number
1061932-2014-01365
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS NOT DISPATCHED TO THE SITE FOR THIS EVENT. CUSTOMER TECHNICAL SPECIALIST (CTS) WAS TOLD BY THE CUSTOMER THAT A LEAK CAME FROM CRACKED TUBING IN THE MIXING MODULE THROUGH FUEL FILTER FF107. THE CUSTOMER WAS ABLE TO REPLACE THE TUBING AND THE INSTRUMENT RAN WITHOUT ANY LEAKS. THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED LESS THAN 5 MLS CLENZ LEAKED BELOW THE COULTER LH 500 HEMATOLOGY ANALYZER. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LAB COAT AT THE TIME OF THE OCCURRENCE. THERE WAS NO EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION WITH THE REPORTED EVENT. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356869 COULTER LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1