FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3882383 · Received June 18, 2014

Report

Report Number
2955842-2014-03770
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
April 8, 2014
Report Date
May 23, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS CONFIRMED THE REPORTED COMPLAINT. THE PITCH CABLE WAS FRAYED AT THE PROXIMAL CLEVIS HUB. FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. THE CLEVIS DID NOT EXHIBIT ANY WEAR. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY WITH BILATERAL SALPINGO-OOPHORECTOMY, BROKEN CABLES WERE IDENTIFIED ON THE PK DISSECTING FORCEPS INSTRUMENT. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
357185 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 N10131218 947

Patients

Seq Age Sex Outcome Treatment
1 39 YR