FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 3882286 · Received June 18, 2014

Report

Report Number
3004209178-2014-85800
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
May 28, 2014
Report Date
May 28, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH DETACHED END CAP. DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WHILE PRIMING INSULIN PUMP, HEARD A NOISE. CUSTOMER FOUND THE BOTTOM PART ALONG WITH THE DRIVE SUPPORT CAP HAD CAME OUT. THE BLOOD GLUCOSE READING WAS 228MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
358164 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551NAB

Patients

Seq Age Sex Outcome Treatment
1 57 YR