MEGA NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2014-03762
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 21, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND ONE GRIP CABLE WAS FRAYED AT THE DISTAL IDLER PULLEY. FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. AN ADDITIONAL OBSERVATION NOT REPORTED WAS THE SAME FRAYED GRIP CABLE WAS DERAILED AT THE DISTAL IDLER PULLEY. THE GRIPS COULD STILL OPEN AND CLOSE BUT MOVEMENT MAY NOT HAVE BEEN PRECISE. THE CABLE DERAILMENT WAS LIKELY DUE TO THE CABLE LOSING CONTACT WITH THE PULLEY DURING WRIST ARTICULATION. THE FLEET ANGLE BETWEEN THE PROXIMAL AND DISTAL PULLEYS MAY HAVE CONTRIBUTED TO THE DERAILMENT. AN ADDITIONAL OBSERVATION NOT REPORTED WERE INDENTATIONS AT THE EDGE OF THE DISTAL PULLEY AND VISIBLE SCRATCHES ON THE SURFACE OF THE PULLEY. FAILURE ANALYSIS CONCLUDED THE INDENTATIONS WERE LIKELY DUE TO MISHANDLING / MISUSE. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY WITH BILATERAL SALPINGO-OOPHORECTOMY PROCEDURE, A BROKEN CABLE WAS IDENTIFIED ON THE MEGA NEEDLE DRIVER INSTRUMENT. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358710 | MEGA NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420194-08 | M10130417 177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |